CO2 Incubators

Back

NU-5831 Direct Heat 7 ft3 (200L) Hypoxic CO2 Incubator with Oxygen Control

The In-VitroCell ES NU-5831 Direct Heat Hypoxic CO2 Incubator can lower oxygen levels in the growth chamber down to 0.5% accurate with in ±0.25% through the use of a Zirconia Ceramic O2 Sensor.

Whatsapp

Overview

Temperature Control

Control Range:

5° above ambient (to a 30° max ambient) to 55°

Set-Point Range: 5°C to 55°C (37.0 Default)
Uniformity: ±0.35°C @ 37°C
Accuracy: ± 0.1°C
Recovery: 0.12°C/min. on Average
Display Resolution: 0.1°C
Door and Perimeter Heater Control Logic: Base Duty Cycle set to chamber temperature Set-Point Proportional 0-100% [manually adjustable]
Temperature Sensor Type: Precision Integrated Circuit

CO2 Control Standard

Range:

0.1 to 20% 5% Default Set Point

Accuracy: ±0.1%
CO2 Recovery: Up to 5% ±0.2% / -0.5 in 5 minutes Average
CO2 Display Resolution: 0.1%
CO2 Control Logic: Fixed Algorithm/Manual, Environmental Adaptable
CO2 Sensor Type: Infrared (NDIR)

Oxygen Display and Control

Control Range:

0.5 -21.0% (ambient) (21.0% Default)

O2 Accuracy: +0.25%
Display Resolution: 0.1%
RH Control Logic: Fixed algorithm/manual environmental adaptable
RH Sensor Type: Zirconia Ceramic